Boonton 4500B RF Peak Power Analyzer
Operation
Table 4-6. Chan Math > Expression > Submenu
Menu Item Function
(Type) Selections (SCPI cmd)
Argument A Pulse, Mod & Stat Selects the first item in the Channel Math
(Mult. Choice) CH1, CH2 expression
Ext Trig 1, Ext Trig 2
Mem1, Mem2 Press the Chan Math > Expression >
Argument A menu key to select either
-
(see below) to complete the mathematical
expression that represents the calculated
waveform you wish to display. Any
combination of waveforms of the same type
can be selected as the arguments of the
expression. Statistical data can only be
compared with statistical data, and likewise
time based data can only be compared with
time based data. Cross comparisons result
in an error condition.
CALCulate:MATH:ARGA {CH1, CH2,
TRIG1, TRIG2, MEM1, MEM2}
Operator Pulse, Mod & Stat Selects the operator for the mathematical
(Mult. Choice) +, -, *, / expression
Press the Chan Math > Expressions >
Operator menu key to select either the plus
or minus sign to indicate whether the
designated waveforms are to be added,
subtracted, multiplied or divided point by
point. Not all operations are valid. See Figure 4-18b.
Result Units below. Chan Math > Expression
CALCulate:MATH:OPERator {CH_PROD, CH_SUM, CH_DIF,
CH_RAT}
Result Units Pulse, Mod & Stat A read only indicator window that describes the validity and form
(Indicator) of the requested operation.
mW the result is the point-by-point sum or difference of the
waveforms.
%Pwr the result is the point-by-point ratio of the waveforms.