Boonton 4500B Peak Power Meter Manual do Utilizador Página 346

  • Descarregar
  • Adicionar aos meus manuais
  • Imprimir
  • Página
    / 372
  • Índice
  • MARCADORES
  • Avaliado. / 5. Com base em avaliações de clientes
Vista de página 345
Boonton 4500B RF Peak Power Analyzer
Maintenance
7-14
7.8 1GHz Calibrator Output Level Calibration
Calibrator 0 dBm Setting
Use the following procedure to calibrate the calibrator 0 dBm output setting:
1. Perform steps 1 10 of the Calibrator 0 dBm Verification procedure in section 7.6
2. If P
o
is between 0.985 mW and 1.015 mW, the output power is within specification. No
calibration is needed. Otherwise, continue to the next step.
3. Enable the UUT cal mode by selecting Spcl > Servicing > Cal Mode > ON.
4. View the UUT fixed cal number by selecting Spcl > Calibrator > Extensions > Fixed Cal.
5. Calculate the new fixed cal (to two decimal places) with the following formula:
N = Fixed Cal [10 × LOG
10
(P
o
)]
Where: N = New fixed cal.
Fixed Cal = Value recorded in Section 7.6 step 14.
P
o
= Value recorded in Section 7.6 step 10.
6. Enter the new fixed cal into the UUT by selecting Spcl > Calibrator > Extensions > Fixed
Cal > N (where N is the number calculated in step 5).
7. Disable UUT cal mode by selecting Spcl > Servicing > Cal Mode > OFF. This step must be
accomplished to set the new EEPROM checksum in the UUT calibrator.
8. Verify the new 0 dBm power output by repeating Section 7.6, steps 1 through 10.
This completes the calibration procedure.
Vista de página 345
1 2 ... 341 342 343 344 345 346 347 348 349 350 351 ... 371 372

Comentários a estes Manuais

Sem comentários