Boonton 4500B Peak Power Meter Manual do Utilizador Página 302

  • Descarregar
  • Adicionar aos meus manuais
  • Imprimir
  • Página
    / 372
  • Índice
  • MARCADORES
  • Avaliado. / 5. Com base em avaliações de clientes
Vista de página 301
Boonton 4500B RF Peak Power Analyzer
Remote Operation
5-102
Table 5-1. SCPI COMMAND SUMMARY (continued)
FETCh[1|2|6|7]:INTERval:PKAVG? Return peak to average power between MK1 and MK2.
FETCh:KEY? Return code of last key depressed
FETCh[1|2|3|4|5|6|7]:MARKer[1|2]:AVERage? Return current reading at the specified marker.
FETCh[1|2|3|4|5|6|7]:MARKer:DELTa? Return difference between MK2 and MK1.
FETCh[1|2|3|4|6|7]:MARKer[1|2]:MAXimum? Return maximum reading at the specified marker
FETCh[1|2|3|4|6|7]:MARKer[1|2]:MINimum? Return minimum reading at the specified marker
FETCh[1|2|3|4|5|6|7]:MARKer:RATio? Return ratio between MK2 and MK1 as a percentage.
FETCh[1|2|3|4|5|6|7]:MARKer:RDELTa? Return difference between MK1 and MK2.
FETCh[1|2|3|4|5|6|7]:MARKer:RRATio? Return ratio between MK1 and MK2 as a percentage.
FETCh[1|2]:TEMPerature:AVERage? Return average sensor internal temperature in degrees C
FETCh[1|2]:TEMPerature:CALTemp? Return sensor calibration temperature in degrees C
FETCh[1|2]:TEMPerature:CURRent? Return current sensor calibration temperature in degrees C
HCOPy:BACKground:COLor Set/return the background color used for printing the screen display. <args> = BLACK,
WHITE, GRAY
INITiate:CONTinuous Set/return state of mode which triggers meas cycles continuously. <arg> = ON, OFF
INITiate[:IMMediate[:ALL]] Set mode which starts a measurement cycle when trigger event occurs
MARKer[1|2]:ASSIgn Assign channel number to marker. <arg> = ACTIVE, CH1, CH2, TRIG1, TRIG2,
CHM, MEM1, MEM2
MARKer:GROUp Sets marker group. <arg> = BOTH, EACH
MARKer:MATH:BOTH Set marker group to both and delta marker, marker math, and units. (Ex: <args> =
RATIO MK1MK2 DBM)
MARKer:MATH:EACH Set marker group to each and delta marker, marker math, and units. (Ex: <args> =
RATIO MK1MK2 DBM)
MARKer[1|2]:POSItion:PERcent Set/return marker position in percent for selected window. <arg> = 0 through 100
MARKer[1|2]:POSItion:PIXel Set/return marker position in pixels for selected window. <arg> = 0 through 500
MARKer[1|2]:POSItion:TIMe Set/return marker time relative to the trigger. Enter time in seconds. (Ex: enter 1ms
as 0.001)