Boonton 4500B Peak Power Meter Manual do Utilizador Página 262

  • Descarregar
  • Adicionar aos meus manuais
  • Imprimir
  • Página
    / 372
  • Índice
  • MARCADORES
  • Avaliado. / 5. Com base em avaliações de clientes
Vista de página 261
Boonton 4500B RF Peak Power Analyzer
Remote Operation
5-62
READ:INTERval:PKAVG?
Description: For the specified channel, return the peak-to-average ratio of the power or voltage between marker
1 and marker 2. The units are dB for logarithmic channel units or percent for linear channel units.
Syntax: READ[1|2|6|7]:INTERval:PKAVG?
Returns: CC, peak-to-average ratio of power or voltage between markers
Where CC is the measurement condition code.
Valid Modes: Pulse and Modulated
READ:MARKer:AVERage?
Description: For the specified channel and marker, return the average power or voltage at the marker. The
units are the same as the specified channel.
Syntax: READ[1|2|3|4|5|6|7]:MARKer[1|2]:AVERage?
Returns: CC, average power or voltage at marker
Where CC is the measurement condition code.
Valid Modes: Pulse and Modulated.
READ:MARKer:DELTa?
Description: For the specified channel return the difference between MK1 and MK2. The units will be the same
as marker units.
Syntax: READ[1|2|3|4|5|6|7]:MARKer:DELTa?
Returns: CC, (MK1 MK2)
Where CC is the measurement condition code.
Valid Modes: All
READ:MARKer:MAXimum?
Description: For the specified channel and marker, return the maximum power or voltage at the marker. The
units are the same as the specified channel.
Syntax: READ[1|2|3|4|6|7]:MARKer[1|2]:MAXimum?
Returns: CC, maximum power or voltage at marker
Where CC is the measurement condition code.
Valid Modes: Pulse and Modulated.
Vista de página 261
1 2 ... 257 258 259 260 261 262 263 264 265 266 267 ... 371 372

Comentários a estes Manuais

Sem comentários