Boonton 4500B RF Peak Power Analyzer
Operation
Table 4-8C. Statistics > Term Settings > Submenu
Menu Item
(Type) Selections Function
Term Time Stat Set terminal run time
(Numeric) Discrete ranges:
Off, 1, 2, 5, 10, 30, 60, Use any of the data entry controls to set a
120, 300, 600, 1800, running time for statistical data collection.
3600 seconds When the time expires, the measurement will
execute the Terminal Action. Select Off to
ignore running time.
TRIGger:CDF:TIMe {0 to 3600}
Term Count Stat Set terminal sample count
(Numeric) Continuous range:
2.0 to 4,096.0 Use any of the data entry controls to set a
Megasamples terminal sample count. When the count is
reached, execute the Terminal Action.
TRIGger:CDF:COUNt <2 to 4096> (M)
Term Action Stat Select the terminal action to execute
(Mult. Choice) Stop, Flush, Decimate
Stop when the terminal time or count is
reached, stop the measurement.
Flush when the terminal time or count is
reached, clear the histograms and buffers
and continue with a new measurement.
Decimate - when the terminal time or count
is reached, divide each histogram bin count
and the total sample count by 2 and continue.
This will provide a continuously updating Fig. 4-26.
distribution function. Statistics > Term Count
menu.
TRIGger:CDF:DECImate {ON, OFF}