Boonton 4500B RF Peak Power Analyzer
Operation
4-92
To recall a previously stored configuration, press the Prgm > Instr Recall menu key. A file system directory display will
appear. Use the data entry controls to select the name of the configuration file you need. Press the Prgm > Instr Recall >
Instrument menu key to restore the recorded setup.
Table 4-39. Prgm > Inst Recall > Menu
Menu Item
(Type) Selections Function
Folder UP Select next higher level folder
(Action)
Folder OPEN Open the selected folder
(Action)
Use any of the data entry controls to move
the selection cursor up and down the list of
files, and select a particular file.
Instrument RECALL Recalls the instrument setup
(Action)
Press the Prgm > Instr Recall > Instrument
menu key to RECALL the stored instrument
configuration file selected above.
Fig. 4-53. Instr > Recall
menu.
Fig. 4-54. Instr > Recall display.