Boonton 4500B RF Peak Power Analyzer
Remote Operation
5.5.14.2 SENSe:HISTogram Data Array Commands
The purpose of the SENS:HISTogram group of commands is to return a power or voltage distribution histogram array. The
HISTogram com
level, and containing the number of samples that have occurred at that power level.
SENSe:HISTogram:COUNt
Description: Set or return the number of histogram bins to be returned as data by SENSe:HISTogram:DATA?
Syntax: SENSe[1|2|3|4|6|7]:HISTogram:COUNt <n>
Argument: <n> = 1 to 16,384
Valid Modes: All
SENSe:HISTogram:DATA[:NEXT]?
Description: Bin values
are 0 to 4,000,000,000.
Syntax: SENSe[1|2|3|4|6|7]:HISTogram:DATA[:NEXT]?
Argument: None, query only
Valid Modes: All
SENSe:HISTogram:INDEX
Description: Set or return the index of the first histogram bin to be returned as data by SENSe:HIST:DATA?
Syntax: SENSe[1|2|3|4|6|7]:HISTogram:INDEX <n>
Argument: <n> = 0 to 16,383 (automatically post-incremented by count)
Valid Modes: All