Boonton 4500B RF Peak Power Analyzer
Operation
4-86
Table 4-33. Calibrator > Pulse (Variable) > Submenu
Menu Item Function
(Type) Selections (SCPI cmd)
Source Int, Ext Selects the source for the calibrator output pulse modulation
(Toggle)
Press the Calibrator > Pulse > menu key to
specify whether the calibrator output pulse is
to be internally or externally generated.
OUTPut:PULSe:SOURce {INT, EXT}
Pulse Width Continuous Range: Set variable mode pulse width
(Numeric) 7 us to 65.535 ms, for
period 65.542 ms, and Use any of the data entry controls to enter
width + 7us period; the pulse width.
Min width increases
with increasing period
above period = 65.544
ms.
See below: OUTPut:PULSe:PERWID
Pulse Period Continuous Range: Set variable mode pulse period
(Numeric) 28 us to 131.077 ms;
Pulse width must be Use any of the data entry controls to enter the
compatible. the pulse period.
OUTPut:PULSe:PERWID <period> <width>
<period> = 28e-6 to 131.07e-3
<width> = 7e-6 to 65.535e-3
Pulse Control Preset, Var Selects the Preset or Variable internal Pulse
(Toggle) modulator
Press the Calibrator > Pulse > Pulse Control Fig. 4-47. Calibrator
menu key to select the Preset method of > Pulse > (Variable
internal pulse modulation control.
See Table 4-31.
OUTPut:PULSe:CTRL {PRESET, VAR}
NOTE
Calibrator pulse modulation can also be supplied by an external pulse generator via the rear-mounted BNC
Calibrator > Pulse > Source menu window. TTL-
level signals connected at this input port will gate the 1 GHz calibration signal on and off. A low level
signal turns the RF output off.