Table 1-2 Model 4500B Performance Specifications (continued)
(Specifications are subject to change without notice)
Trigger Delay Resolution: 0.02 divisions
Trigger Hold-off Range: 0.0 1.0 sec
Trigger Hold-off Resolution: 10 nsec
Trigger Mode: Normal, Auto, Auto pk-to-pk, Free Run
B Trigger Mode: A only, B Delay-by-time, B Delay-by-events
B Trigger Source: Chan 1, Chan 2, Ext Trig 1, Ext Trig 2
B Trigger Slope: + or
B Trigger Events Counter Range: 1 to 999,999 events
B Trigger Time Delay Range: 0.0 1.0 sec in 10 nsec increments
B Trigger Time Delay Resolution: 10 nsec
Internal Trigger Level Range: -39.9 to +20 dBm (sensor dependent)
External Trigger Level Range: ±5 volts, ±50 volts
External Trigger Input: 1 Megohm in parallel with 70 pF, or 50 ohm, DC Coupled
Trigger Output (option 06): One TTL output per channel via rear-panel BNC connector(s). Output
indicates state of channel trigger system. Latency <30ns, minimum
pulse width 20ns, max pulse frequency 20MHz (trigger source limited).
CALIBRATION SOURCE
Operating Modes: CW, Internal Pulse or External Pulse
Frequency: 1.024 GHz ± 0.01%
Level Range: -50 to +20 dBm
Resolution: 0.1 dB
Output VSWR: 1.20 maximum
CW Absolute Accuracy: ±0.065 dB (±1.5%) at 0 dBm
CW Accuracy vs level: add ±0.03 dB per 5 dB increment from 0 dBm
Preset Internal Pulse Period: 100 usec, 1 msec or 10 msec
Preset Internal Pulse Duty Cycle: 10% to 90% in 10% increments
Variable Pulse On Time: 7 usec to 65.535 msec in 1 usec steps
Variable Pulse Period: 28 usec to 131.072 msec in 2 usec steps
Off-time limits - within 7 usec to 65.535 msec
Pulse Polarity; + or
RF Connector: Precision Type N
External Pulse Input: Rear panel BNC, TTL level compatible
Auto-calibration: The Calibrator is used to automatically generate linearity calibration
data for peak power sensors. It is also provides test signals.