Boonton 4500B RF Peak Power Analyzer
Operation
AUTOMATIC MEASUREMENTS – STATISTICAL MODE (Opt 10 only)
In the Statistical mode automatic measurements are made Top Level > Measurement >
Stop selection is made, the latest value is displayed. Press the Top Level > Display > CLEAR menu key to clear previous
results. Press the Top Level > Measurement >
until a terminate condition is met. The automatic measurements will continuously update with the results of the accumulating
data. The peak-to-average reading is not measured on External Trigger channels. The peak-to-average and dynamic range
results are always displayed in dB units. The average, peak and minimum results will have the same units as the associated
channel.
Table 4-61. Statistical Mode – Automatic Measurements I
CH1&2 CH 1&2
Name (Power Units) (Voltage Units)
Avg Average power Average voltage (converted)
Peak Peak power Peak voltage
Min Minimum power Minimum voltage
Pk/Avg Peak-to-average ratio (dB) Peak-to-average ratio (dB)
Dyn Rng Dynamic range (dB) Dynamic range (dB)
Table 4-62. Statistical Mode – Automatic Measurements II
EXT TRIG 1&2
Name (Voltage Units)
Avg Average voltage
Peak Peak voltage
Min Minimum voltage
Pk/Avg ----
Dyn Rng Peak-to-peak voltage