Boonton 4500B RF Peak Power Analyzer
Operation
Table 4-32. Calibrator > Pulse (Preset) > Submenu (continued)
Menu Item Function
(Type) Selections (SCPI cmd)
Duty Cycle Discrete Range: Selects the duty cycle for calibrator’s preset internal pulse modulation
(Numeric) + Polarity:
10, 20, 30, 40, 50 % Press the Calibrator > Pulse > Duty Cycle menu key to step through
Polarity: the range of available duty cycles. This control has no effect on
90, 80, 70, 10, 50 % External or Internal Variable pulse modulation.
OUTPut:PULSe:DCYCle {10, 20, 30, 40, 50}
Pulse Period Discrete Range Selects the period of the calibrator’s preset pulse modulation
(Numeric) 100 us, 1 ms, 10 ms
Press the Calibrator > Pulse > Pulse Period
menu key to step through the range of
available pulse periods.
OUTPut:PULSe:PERiod {0.1, 1, 10}
Pulse Control Preset, Var Selects the Preset or Variable internal Pulse
(Toggle) modulator
Press the Calibrator > Pulse > Pulse Control
menu key to select the Variable method of
internal pulse modulation control.
See Table 4-32.
OUTPut:PULSe:CTRL {PRESET, VAR}
Fig. 4-46. Calibrator >
Pulse > (Preset)