Boonton 4500B RF Peak Power Analyzer
Remote Operation
5.5.4 DISPlay Subsystem
The DISPlay group of commands is used to control the selection and presentation of textual, graphical and TRACe
measurements.
DISPlay:ACTive
Description: Set or return the active channel. The active channel trace is drawn on top of all other traces. The
the trace area if status display is selected.
Syntax: DISPlay:ACTive <asc>
Argument: <asc> = {CH1, CH2, TRIG1, TRIG2, CHM, MEM1, MEM2}
Valid Modes: All
DISPlay:BACKlight:BRIGhtness
Description: Set or return the backlight brightness for the LCD display in percent.
Syntax: DISPlay:BACKlight:BRIGhtness <n>
Argument: <n> = 10 to 100
Valid Modes: All
DISPlay:BRIGhtness:COLor
Description: Set or return the color item to be set by the RGB commands that follow.
Syntax: DISPlay:BRIGhtness COLor <asc>
Argument: <asc> = see Color Item Table below
Color Item Table {
BACKGROUND GRID TIME_MARKERS
CH1 CH2 TRIG1
TRIG2 CH_MATH MEM1
MEM2 REF_LINES STATUS_MSG
ERROR_MSG PATH_MSG PRIORITY_MSG
BOX_LOW BOX_HIGH LABEL_BACK
MENU_LABEL DATA_BACK MENU_DATA
DATA_LOW TEXT}
Valid Modes: All