Boonton 4500B RF Peak Power Analyzer
Remote Operation
5-17
5.5.4 DISPlay Subsystem
The DISPlay group of commands is used to control the selection and presentation of textual, graphical and TRACe
measurements.
DISPlay:ACTive
Description: Set or return the active channel. The active channel trace is drawn on top of all other traces. The

the trace area if status display is selected.
Syntax: DISPlay:ACTive <asc>
Argument: <asc> = {CH1, CH2, TRIG1, TRIG2, CHM, MEM1, MEM2}
Valid Modes: All
DISPlay:BACKlight:BRIGhtness
Description: Set or return the backlight brightness for the LCD display in percent.
Syntax: DISPlay:BACKlight:BRIGhtness <n>
Argument: <n> = 10 to 100
Valid Modes: All
DISPlay:BRIGhtness:COLor
Description: Set or return the color item to be set by the RGB commands that follow.
Syntax: DISPlay:BRIGhtness COLor <asc>
Argument: <asc> = see Color Item Table below
Color Item Table {
BACKGROUND GRID TIME_MARKERS
CH1 CH2 TRIG1
TRIG2 CH_MATH MEM1
MEM2 REF_LINES STATUS_MSG
ERROR_MSG PATH_MSG PRIORITY_MSG
BOX_LOW BOX_HIGH LABEL_BACK
MENU_LABEL DATA_BACK MENU_DATA
DATA_LOW TEXT}
Valid Modes: All