Boonton 4500B RF Peak Power Analyzer
Operation
4-83
4.16 CAL Key and Calibrator > Menu
With the Calibrator > menu (Figure 4-45), you can turn on the calibrator output and adjust the calibrator signal parameters.
The items in the Calibrator> menu and its two submenus are presented in the following figures and tables:
Menu or Submenu Figure Table
Calibrator > 4-45 4-31
Calibrator > Pulse (Preset) 4-46 4-32
Calibrator > Pulse (Variable) 4-47 4-33
Calibrator > Extensions 4-48 4-34
Table 4-31. Calibrator > Menu
Menu Item Function
(Type) Selections (SCPI cmd)
Cal Output Off, On Controls the calibrator RF output.
(Toggle)
Press the Calibrator > Cal Output menu key


OUTPut:SIGnal {ON, OFF}
Set Level Continuous Range: Sets the calibrator output level.
(Numeric) -50.0 to +20.0 dBm
Press the Calibrator > Set Level menu key
to activate this function. Use the data entry
controls to adjust the calibrator output
power to the desired level.
OUTPut:LEVel <-50.0 to 20.0>
Max Power Continuous Range: Establishes the upper limit for the calibrator
(Numeric) -50.0 to +20.0 dBm output power.
Press the Calibrator > Max Power menu key
to activate this function. Use the data entry
controls to adjust the upper limit of calibrator
output power.
An error message will be generated if you
attempt to set the Calibrator > Max Power
value lower than Set Level, or if you attempt Fig. 4-45. Calibrator >
menu.