Remote Operation
DISPlay:PARAmeter:MODE
Description: Set or return the type of content displayed in the parameter area above the trace box. The STAT
(status) command causes a fixed set of nine status parameters for the active channel to be
displayed. The MEAS command causes a group of nine user selected auto-measure and fixed
parameters from any measurement channel to be displayed.
Syntax: DISPlay:PARAmeter:MODE <asc>
Argument: <asc> = {MEAS, STAT}
Valid Modes: Pulse Mode only. Note: Modulated mode does not display Status and Measurement parameters.
Statistical mode displays a fixed combination of status and auto-measure parameters.
DISPlay:CHANnel:BOTtom:LEFT
Description: Set or return the channel for the auto-measure or fixed parameter displayed in the Bottom row,
Left column of the nine element parameter field above the trace box.
DISPlay:PARAmeter:MODE must be set to MEAS.
Syntax: DISPlay:CHANnel:BOTtom:LEFT <asc>
Argument: <asc> = {CH1, CH2, TRIG1, TRIG2, CHM, MEM1, MEM2}
Valid Modes: Pulse Mode only. Note: Modulated mode does not display Status and Measurement parameters.
Statistical mode displays a fixed combination of status and auto-measure parameters.
DISPlay:CHANnel:BOTtom:MIDdle
Description: Set or return the channel for the auto-measure or fixed parameter displayed in the Bottom row,
Middle column of the nine element parameter field above the trace box.
DISPlay:PARAmeter:MODE must be set to MEAS.
Syntax: DISPlay:CHANnel:BOTtom: MIDdle <asc>
Argument: <asc> = {CH1, CH2, TRIG1, TRIG2, CHM, MEM1, MEM2}
Valid Modes: Pulse Mode only. Note: Modulated mode does not display Status and Measurement parameters.
Statistical mode displays a fixed combination of status and auto-measure parameters.
DISPlay:CHANnel:BOTtom:RIGHt
Description: Set or return the channel for the auto-measure or fixed parameter displayed in the Bottom row,
Right column of the nine element parameter field above the trace box.
DISPlay:PARAmeter:MODE must be set to MEAS.
Syntax: DISPlay:CHANnel:BOTtom: RIGHt <asc>
Argument: <asc> = {CH1, CH2, TRIG1, TRIG2, CHM, MEM1, MEM2}
Valid Modes: Pulse Mode only. Note: Modulated mode does not display Status and Measurement parameters.
Statistical mode displays a fixed combination of status and auto-measure parameters.