Boonton 4540 Peak Power Meter User Manual Manual do Utilizador Página 1

Consulte online ou descarregue Manual do Utilizador para Instrumentos de medida Boonton 4540 Peak Power Meter User Manual. Boonton 4540 Peak Power Meter User Manual User Manual Manual do Utilizador

  • Descarregar
  • Adicionar aos meus manuais
  • Imprimir
  • Página
    / 270
  • Índice
  • MARCADORES
  • Avaliado. / 5. Com base em avaliações de clientes

Resumo do Conteúdo

Página 1

Wireless Telecom Group 25 EASTMANS ROAD, PARSIPPANY, NJ 07054 Telephone: 973-386-9696 Fax: 973-386-9191 Email: [email protected] Web: www

Página 2

Boonton 4540 Series RF Power Meter Contents x MEMory:SNSR:LFLIN? ... 5

Página 3 - SAFETY SUMMARY

Boonton 4540 Series RF Power Meter Operation 4-52 4.5.31 Display > Graph Header Menu The Graph header menu is used to customize the text fields a

Página 4

Boonton 4540 Series RF Power Meter Operation 4-53 Table 4-1 Field Parameter Selection List Field Parameter Description Pulse Width Automatic Pulse

Página 5

Boonton 4540 Series RF Power Meter Operation 4-54 4.5.32 Display > Text Mode Menu The Text Mode menu can be used to configure the Modulated text

Página 6

Boonton 4540 Series RF Power Meter Operation 4-55 4.5.33 Display > Text Mode > Ch1|2 Options Menu The Ch1|2 Options menu provides methods for

Página 7

Boonton 4540 Series RF Power Meter Operation 4-56 Main Pg 2 > Display > Text Mode > Ch1|2 Options > Field 2 Chan Description: Select chan

Página 8

Boonton 4540 Series RF Power Meter Operation 4-57 4.5.34 Display > Backlight Menu The Backlight menu provides methods for adjusting backlight and

Página 9

Boonton 4540 Series RF Power Meter Operation 4-58 4.5.35 System Menu The System menu provides access to system configuration, information, and contro

Página 10 - Contents

Boonton 4540 Series RF Power Meter Operation 4-59 4.5.36 System > I/O Config Menu The I/O Config menu provides access to set GPIB address and con

Página 11

Boonton 4540 Series RF Power Meter Operation 4-60 4.5.37 System > I/O Config > Ethernet Menu The Ethernet menu provides access to configuratio

Página 12

Boonton 4540 Series RF Power Meter Operation 4-61 4.5.38 System > I/O Config > Ethernet > Extensions Menu The Ethernet Extensions menu cont

Página 13

Boonton 4540 Series RF Power Meter Contents xi READ:MARKer:CURsor:POWer? ... 5-50 READ

Página 14

Boonton 4540 Series RF Power Meter Operation 4-62 Main Pg 2 > System > I/O Config > Ethernet > Extensions > MAC Address Description:

Página 15 - 1. General Information

Boonton 4540 Series RF Power Meter Operation 4-63 4.5.39 System > I/O Config > Multi IO Menu This menu selects the signal to be sent to the Mu

Página 16 - General Information

Boonton 4540 Series RF Power Meter Operation 4-64 4.5.40 System > I/O Config > Multi IO > IO Mode Status Out Menu This menu selects a TTL c

Página 17 - 1.3 Features

Boonton 4540 Series RF Power Meter Operation 4-65 4.5.41 System > I/O Config > Multi IO > IO Mode Rec Out Menu This menu selects an analog

Página 18

Boonton 4540 Series RF Power Meter Operation 4-66 Main Pg 2 > System > I/O Config > Multi IO > Scaling (Rec Out) Description: Sele

Página 19 - 1.4 Accessories

Boonton 4540 Series RF Power Meter Operation 4-67 4.5.43 System > I/O Config > Multi IO > IO Mode Trig Out Menu This menu selects trigger r

Página 20 - 1.6 Specifications

Boonton 4540 Series RF Power Meter Operation 4-68 4.5.44 System > I/O Config > Multi IO > IO Mode Volt Out Menu This menu selects a voltage

Página 21

Boonton 4540 Series RF Power Meter Operation 4-69 4.5.45 System > Calibrator Menu (Internal 50 MHz) The Calibrator Menu is used to select and con

Página 22

Boonton 4540 Series RF Power Meter Operation 4-70 4.5.46 System > Calibrator Menu (External 1 GHz - optional) The Calibrator Menu is used to secl

Página 23

Boonton 4540 Series RF Power Meter Operation 4-71 Main Pg 2 > System > Calibrator > Status Description: This menu item is not implemented a

Página 24

Boonton 4540 Series RF Power Meter Contents xii STATus:QUEStionable:EVENt? ... 5-6

Página 25 - 2. Installation

Boonton 4540 Series RF Power Meter Operation 4-72 4.5.47 System > Calibrator > Pulse - Preset Menu (Ext 1 GHz - opt.) The pulse menu is speci

Página 26 - 2.3 Connections

Boonton 4540 Series RF Power Meter Operation 4-73 4.5.48 System > Calibrator > Pulse - Variable Menu (Ext 1 GHz - opt.) The pulse menu is spe

Página 27 - 2.4 Preliminary Check

Boonton 4540 Series RF Power Meter Operation 4-74 FFF Fig. 4-62.

Página 28

Boonton 4540 Series RF Power Meter Operation 4-75 4.5.49 System > Reports Menu System Reports are text displays that provide additional informat

Página 29 - 3. Getting Started

Boonton 4540 Series RF Power Meter Operation 4-76 Main Pg 2 > System > Reports > Info The Hardware Info Report shows temperature, power sup

Página 30 - Getting Started

Boonton 4540 Series RF Power Meter Operation 4-77 Main Pg 2 > System > Reports > Remote Cmds The Remote Commands Report is a list of all 454

Página 31

Boonton 4540 Series RF Power Meter Operation 4-78 4.5.50 System > Servicing Menu This menu accesses various service related functions. Main Pg

Página 32

Boonton 4540 Series RF Power Meter Operation 4-79 4.5.51 System > Servicing > Security Menu This menu provides a means to erase sensitive stor

Página 33

Boonton 4540 Series RF Power Meter Operation 4-80 Main Pg 2 > System > Servicing > Security > Full Sanitize Description: Select

Página 34 - 3.3 LCD Monitor Display

Boonton 4540 Series RF Power Meter Operation 4-81 4.5.52 System > Sensor Data Menu This menu provides information about the characteristics and c

Página 35

Boonton 4540 Series RF Power Meter Contents xiii TRIGger:POSition ...

Página 36

Boonton 4540 Series RF Power Meter Operation 4-82 Main Pg 2 > System > Sensor Data > Sensor 2 Info Description: See a typical CW sensor re

Página 37

Boonton 4540 Series RF Power Meter Operation 4-83 4.5.53 Setup Menu This menu provides for system preset with default parameters, pre-defined parame

Página 38 - Caution

Boonton 4540 Series RF Power Meter Operation 4-84 4.5.54 Setup > Preset Menu This menu provides presets for widely used communications system sig

Página 39

Boonton 4540 Series RF Power Meter Operation 4-85 4.5.55 Setup > User Presets Menu Main Pg 2 > Setup > User Preset > Preset Name Descr

Página 40

Boonton 4540 Series RF Power Meter Operation 4-86 4.6 Cal/Zero Key Menu Pressing thefront panel Cal/Zero key wil

Página 41

Boonton 4540 Series RF Power Meter Remote Operation 5-1 5. Remote Operation 5.1 GPIB Configuration The 4540 GPIB interface is configured using the

Página 42 - STEP PROCEDURE

Boonton 4540 Series RF Power Meter Remote Operation 5-2 5.4 SCPI Language The 4540 Series instruments may be remotely controlled using comm

Página 43

Boonton 4540 Series RF Power Meter Remote Operation 5-3 5.4.5 Command Arguments and Queries Many commands require arguments. In this case, the entir

Página 44

Boonton 4540 Series RF Power Meter Remote Operation 5-4 :CALCulate:STATe? Query the state of measurement channel 1 (default chan #) :CALC:STAT? Quer

Página 45 - 3.6 Making Measurements

Boonton 4540 Series RF Power Meter Remote Operation 5-5 5.5 Basic Measurement Information The easiest way to obtain a reading is by use of the MEASu

Página 46

Boonton 4540 Series RF Power Meter Contents xiv 9.3 Optional Configurations ...

Página 47

Boonton 4540 Series RF Power Meter Remote Operation 5-6 5.6 SCPI Command Reference This section contains a list of all SCPI remote commands accepted

Página 48

Boonton 4540 Series RF Power Meter Remote Operation 5-7 *ESR? Description: Return the current value of the Standard Event Status Register, then cl

Página 49 - 4. Operation

Boonton 4540 Series RF Power Meter Remote Operation 5-8 *OPC? Description: This command examines the OPC (Operation Complete) status fla

Página 50 - 4.3 Numerical Data Entry

Boonton 4540 Series RF Power Meter Remote Operation 5-9 *STB? Description: Return the current value of the Status Byte register. This register ha

Página 51 - 4.4 Sensor Calibration

Boonton 4540 Series RF Power Meter Remote Operation 5-10 *WAI Description: Wait command. This command insures sequential, non-overlapped executio

Página 52 - 4.5 Menu Reference

Boonton 4540 Series RF Power Meter Remote Operation 5-11 5.6.2 CALCulate Subsystem The CALCulate group of the command subsystem is used to configure

Página 53

Boonton 4540 Series RF Power Meter Remote Operation 5-12 CALCulate:LIMit:UPPer[:POWer] Description: Set or return the upper limit power level for

Página 54

Boonton 4540 Series RF Power Meter Remote Operation 5-13 CALCulate:MATH Description: Set or return the signal source or sources combined in an ari

Página 55

Boonton 4540 Series RF Power Meter Remote Operation 5-14 CALCulate:REFerence:COLLect Description: For the selected channel, make the current powe

Página 56

Boonton 4540 Series RF Power Meter Remote Operation 5-15 CALCulate:STATe Description: Set or return the measurement state of the selected channel.

Página 57

Boonton 4540 Series RF Power Meter General Information 1-1 1. General Information This instruction manual provides you with the information you need

Página 58

Boonton 4540 Series RF Power Meter Remote Operation 5-16 5.6.3 CALibration Subsystem The CALibration group of commands is used to control automatic z

Página 59

Boonton 4540 Series RF Power Meter Remote Operation 5-17 CALibration[:INTernal]:AUTOcal Description: Performs a multi-point sensor gain calibratio

Página 60

Boonton 4540 Series RF Power Meter Remote Operation 5-18 5.6.4 DISPlay Subsystem The DISPlay group of commands is used to control the selec

Página 61

Boonton 4540 Series RF Power Meter Remote Operation 5-19 DISPlay:MODUlated:TIMEBASE Description: Set or return the Modulated Mode timebase in s

Página 62

Boonton 4540 Series RF Power Meter Remote Operation 5-20 DISPlay:PULSe:TSPAN (Note: This command is supplied for compatibility with the Model

Página 63

Boonton 4540 Series RF Power Meter Remote Operation 5-21 DISPlay:[TEXt:]LIN:RESolution Description: Set or return the display resolution for

Página 64

Boonton 4540 Series RF Power Meter Remote Operation 5-22 DISPlay:TRACe:VCENTer Description: Set or return the power or voltage level of the horizo

Página 65

Boonton 4540 Series RF Power Meter Remote Operation 5-23 5.6.5 FETCh Queries The FETCh? group of queries is used to return specific measurement data

Página 66

Boonton 4540 Series RF Power Meter Remote Operation 5-24 FETCh:ARRay:AMEAsure:TIMe? Description: Returns an array of the current automatic timing

Página 67

Boonton 4540 Series RF Power Meter Remote Operation 5-25 FETCh:ARRay:MARKer:POWer? Description: Returns an array of the current marker measurement

Página 68

Boonton 4540 Series RF Power Meter General Information 1-2 1.2 Description The 4540 Series RF Power Meter is the latest generation of BOON

Página 69

Boonton 4540 Series RF Power Meter Remote Operation 5-26 FETCh:INTERval:AVERage? Description: For the specified channel, return the average power

Página 70

Boonton 4540 Series RF Power Meter Remote Operation 5-27 FETCh:INTERval:MINimum? Description: For the specified channel, return the minimum insta

Página 71

Boonton 4540 Series RF Power Meter Remote Operation 5-28 FETCh:MARKer:CURsor:PERcent? Description: Return the CCDF cursor y-axis position in

Página 72

Boonton 4540 Series RF Power Meter Remote Operation 5-29 FETCh:MARKer:MINimum? Description: For the specified channel and marker, return the minim

Página 73

Boonton 4540 Series RF Power Meter Remote Operation 5-30 FETCh:TEMPerature:AVERage? Description: Return the averaged internal temperature of

Página 74

Boonton 4540 Series RF Power Meter Remote Operation 5-31 5.6.6 INITiate and ABORt Commands The purpose of the INITiate group of commands is to start

Página 75

Boonton 4540 Series RF Power Meter Remote Operation 5-32 5.6.7 MARKer Subsystem The MARKer group of commands is used to configure and locate measurem

Página 76

Boonton 4540 Series RF Power Meter Remote Operation 5-33 MARKer:POSItion:TIMe Description: Set or return the time (x-axis-position) of the selecte

Página 77

Boonton 4540 Series RF Power Meter Remote Operation 5-34 5.6.8 MEASure Queries The MEASure group of commands is used to acquire data using a set of h

Página 78

Boonton 4540 Series RF Power Meter Remote Operation 5-35 5.6.9 MEMory Subsystem The MEMory group of commands is used to save and recall instrument op

Página 79

Boonton 4540 Series RF Power Meter General Information 1-3 1.3 Features Software. A 32-bit control computer running a real-time operating system

Página 80

Boonton 4540 Series RF Power Meter Remote Operation 5-36 MEMory:SNSR:INFO? Description: Return the sensor ID and parameter data. Syntax: MEMor

Página 81

Boonton 4540 Series RF Power Meter Remote Operation 5-37 MEMory:SYS:LOAD Description: Recall a previously stored file image of an instrument setup

Página 82

Boonton 4540 Series RF Power Meter Remote Operation 5-38 5.6.10 OUTPut Subsystem The OUTPut group of commands is used to control various outputs of t

Página 83

Boonton 4540 Series RF Power Meter Remote Operation 5-39 OUTPut:EXTernal:PRESent? Description: Return true if the Model 2530 external 1 GHz calib

Página 84

Boonton 4540 Series RF Power Meter Remote Operation 5-40 OUTPut:EXTernal:PULse:PERWID Description: Set or return the pulse period <arg1> and

Página 85

Boonton 4540 Series RF Power Meter Remote Operation 5-41 OUTPut:INTernal:SIGNal Description: Set or return the on/off state of the internal 50 M

Página 86

Boonton 4540 Series RF Power Meter Remote Operation 5-42 OUTPut:[MIO:]RECOroderMIN Description: Set or return the Multi IO recorder output

Página 87

Boonton 4540 Series RF Power Meter Remote Operation 5-43 OUTPut:MIO:STATus:SETTing Description: Set/return Multi IO status out setting. Below is

Página 88

Boonton 4540 Series RF Power Meter Remote Operation 5-44 OUTPut:MIO:TRIGout:SOURce Description: Set/return Multi IO source for trigger out mode.

Página 89

Boonton 4540 Series RF Power Meter Remote Operation 5-45 5.6.11 READ Queries The purpose of the READ? group of queries is to initiate a measurement c

Página 90

Boonton 4540 Series RF Power Meter General Information 1-4 Dual Independent Channels. When equipped with the optional second measurement channel, t

Página 91

Boonton 4540 Series RF Power Meter Remote Operation 5-46 READ:ARRay:AMEAsure:TIMe? Description: Returns an array of the current automatic timing m

Página 92

Boonton 4540 Series RF Power Meter Remote Operation 5-47 READ:ARRay:MARKer:POWer? Description: Returns an array of the current marker measurement

Página 93

Boonton 4540 Series RF Power Meter Remote Operation 5-48 READ:INTERval:MINFilt? Description: For the specified channel, return the minimum

Página 94

Boonton 4540 Series RF Power Meter Remote Operation 5-49 READ:MARKer:AVERage? Description: For the specified channel and marker, return the averag

Página 95

Boonton 4540 Series RF Power Meter Remote Operation 5-50 READ:MARKer:CURsor:PERcent? Description: Return the CCDF cursor y-axis position in

Página 96

Boonton 4540 Series RF Power Meter Remote Operation 5-51 READ:MARKer:RDELTa? Description: For the specified channel return the difference between

Página 97

Boonton 4540 Series RF Power Meter Remote Operation 5-52 5.6.12 SENSe Subsystem The purpose of the SENSe command subsystem is to directly configure d

Página 98

Boonton 4540 Series RF Power Meter Remote Operation 5-53 SENSe:ARRay:CALTemp? Description: Returns both channels’ sensor calibration temperatu

Página 99

Boonton 4540 Series RF Power Meter Remote Operation 5-54 SENSe:CORRection:FREQuency Description: Set or return the RF frequency for the cu

Página 100 - Operation

Boonton 4540 Series RF Power Meter Remote Operation 5-55 SENSe:FILTer:STATe Description: Set or return the current setting of the integration filt

Página 101

Boonton 4540 Series RF Power Meter General Information 1-5 1.4 Accessories Optional 4540 accessories that can be ordered from Boonton Electronics.

Página 102

Boonton 4540 Series RF Power Meter Remote Operation 5-56 SENSe:PULSe:ENDGT Description: Set or return the point on a pulse, which is used to defin

Página 103

Boonton 4540 Series RF Power Meter Remote Operation 5-57 SENSe:PULSe:STARTGT Description: Set or return the point on a pulse, which is used to def

Página 104

Boonton 4540 Series RF Power Meter Remote Operation 5-58 5.6.13 SENSe:MBUF Data Array Commands The MBUF commands (which are grouped under the SENSe s

Página 105

Boonton 4540 Series RF Power Meter Remote Operation 5-59 SENSe:MBUF:SIZe Description: Set or return the size of the measurement buffer.

Página 106

Boonton 4540 Series RF Power Meter Remote Operation 5-60 SENSe:MBUF:POSition Description: Return the current acquisition (write) position into the

Página 107

Boonton 4540 Series RF Power Meter Remote Operation 5-61 SENSe:MBUF:MEASurement Description: Set or return the desired measurement value to be

Página 108

Boonton 4540 Series RF Power Meter Remote Operation 5-62 5.6.14 STATus Commands The STATus command subsystem enables you to control the SCPI

Página 109

Boonton 4540 Series RF Power Meter Remote Operation 5-63 STATus:DEVice:EVENt? Description: Returns the current contents of the Device Event regist

Página 110

Boonton 4540 Series RF Power Meter Remote Operation 5-64 STATus:OPERation:CONDition? Description: Return the current value of the Operation Condit

Página 111

Boonton 4540 Series RF Power Meter Remote Operation 5-65 STATus:OPERation:NTRansition Description: Set or return the value of the negative transit

Página 112

Boonton 4540 Series RF Power Meter Contents ii INSTRUCTION MANUAL, 4540 SERIES RF POWER METER Revision date 10/04/2012 © Copyri

Página 113

Boonton 4540 Series RF Power Meter General Information 1-6 1.6 Specifications Performance specifications for the 4540 Series are listed in Table 1-2

Página 114

Boonton 4540 Series RF Power Meter Remote Operation 5-66 STATus:QUEStionable:CONDition? Description: Return the current value of the Questionable

Página 115

Boonton 4540 Series RF Power Meter Remote Operation 5-67 STATus:QUEStionable:EVENt? Description: Returns the current contents of the Questionable

Página 116

Boonton 4540 Series RF Power Meter Remote Operation 5-68 STATus:QUEStionable:CALibration:CONDition? Description: Return the current value of the Q

Página 117

Boonton 4540 Series RF Power Meter Remote Operation 5-69 STATus:QUEStionable:CALibration:NTRansition Description: Set or return the value of the neg

Página 118

Boonton 4540 Series RF Power Meter Remote Operation 5-70 5.6.15 SYSTem Subsystem The SYSTem group of commands is used to control system-level functio

Página 119

Boonton 4540 Series RF Power Meter Remote Operation 5-71 SYSTem:COMMunicate:LAN:ADDRess Description: Set or return the IP address for the Ethernet

Página 120

Boonton 4540 Series RF Power Meter Remote Operation 5-72 SYSTem:COMMunicate:LAN:MAC? Description: Return the instrument’s Ethernet MAC address.

Página 121

Boonton 4540 Series RF Power Meter Remote Operation 5-73 SYSTem:DATE:MONTH Description: Set or return the month of the real time clock. Syntax:

Página 122

Boonton 4540 Series RF Power Meter Remote Operation 5-74 SYSTem:ERRor:CODE? Description: Returns the next queued error code number. Note t

Página 123

Boonton 4540 Series RF Power Meter Remote Operation 5-75 SYSTem:TIMe:HOUR Description: Set or return the hour of the real time clock. Syntax:

Página 124

Boonton 4540 Series RF Power Meter General Information 1-7 Table 1-2 4540 Series Performance Specifications (continued) (Specifications are subje

Página 125

Boonton 4540 Series RF Power Meter Remote Operation 5-76 5.6.16 TRACe Data Array Commands The TRACe group of commands is used to control the output o

Página 126

Boonton 4540 Series RF Power Meter Remote Operation 5-77 TRACe:COUNt Description: Set or return the number of trace points, which will be returned

Página 127

Boonton 4540 Series RF Power Meter Remote Operation 5-78 5.6.17 TRIGger Subsystem The TRIGger group of commands is used to control synchronization of

Página 128

Boonton 4540 Series RF Power Meter Remote Operation 5-79 TRIGger:DELay Description: Set or return the trigger delay time in seconds with respect t

Página 129

Boonton 4540 Series RF Power Meter Remote Operation 5-80 TRIGger:MODe Description: Set or return the trigger mode for synchronizing data acquisit

Página 130

Boonton 4540 Series RF Power Meter Remote Operation 5-81 TRIGger:SOURce Description: Set or return the trigger source used for synchroniz

Página 131

Boonton 4540 Series RF Power Meter Remote Operation 5-82 5.6.18 SCPI Command Summary Table 5-1. SCPI COMMAND SUMMARY *CLS Clear Status Command *

Página 132

Boonton 4540 Series RF Power Meter Remote Operation 5-83 Table 5-1. SCPI COMMAND SUMMARY (continued) CALCulate[1|2]:PKHLD Set/return state of the p

Página 133

Boonton 4540 Series RF Power Meter Remote Operation 5-84 Table 5-1. SCPI COMMAND SUMMARY (continued) DISPlay:SCREensaver:STATe Set/ret

Página 134 - 4.6 Cal/Zero Key Menu

Boonton 4540 Series RF Power Meter Remote Operation 5-85 5-1. SCPI COMMAND SUMMARY (continued) FETCh[1|2]:INTERval:PKAVG? Return peak to avera

Página 135 - 5. Remote Operation

Boonton 4540 Series RF Power Meter General Information 1-8 Table 1-2 4540 Series Performance Specifications (continued) (Specifications are subje

Página 136 - 5.4 SCPI Language

Boonton 4540 Series RF Power Meter Remote Operation 5-86 Table 5-1. SCPI COMMAND SUMMARY (continued) MEMory[1|2]:SNSR:CFFAST? Return the sensor hig

Página 137

Boonton 4540 Series RF Power Meter Remote Operation 5-87 Table 5-1. SCPI COMMAND SUMMARY (continued) OUTPut:EXTernal:SIGNal Set or return the on/of

Página 138

Boonton 4540 Series RF Power Meter Remote Operation 5-88 Table 5-1. SCPI COMMAND SUMMARY (continued) READ[1|2]:ARRay:AMEAsure:POWer? Return new Pp

Página 139

Boonton 4540 Series RF Power Meter Remote Operation 5-89 Table 5-1. SCPI COMMAND SUMMARY (continued) SENSe[1|2]:CORRection:CALFactor Set co

Página 140 - 5.6 SCPI Command Reference

Boonton 4540 Series RF Power Meter Remote Operation 5-90 Table 5-1. SCPI COMMAND SUMMARY (continued) SENSe:MBUF:COUNt Set/return number of buffer p

Página 141

Boonton 4540 Series RF Power Meter Remote Operation 5-91 Table 5-1. SCPI COMMAND SUMMARY (continued) SYSTem:AUTOSET Perfor

Página 142

Boonton 4540 Series RF Power Meter Remote Operation 5-92 Table 5-1. SCPI COMMAND SUMMARY (continued) SYSTem:TIMe:MINUTE Set/ge

Página 143

Boonton 4540 Series RF Power Meter Remote Operation 5-93 This page intentionally left blank.

Página 145

Boonton 4540 Series RF Power Meter Application Notes 6-1 6. Application Notes This section provides supplementary material to enhance your knowledge

Página 146

Boonton 4540 Series RF Power Meter General Information 1-9 Table 1-2 4540 Series Performance Specifications (continued) (Specifications are subje

Página 147

Boonton 4540 Series RF Power Meter Application Notes 6-2 Figure 6-1. Pulsed RF Signal Figure 6-2. Distorted Pulse Signal

Página 148

Boonton 4540 Series RF Power Meter Application Notes 6-3 6.1.2 Diode Detection Wideband diode detectors are the dominant power sensing device used

Página 149

Boonton 4540 Series RF Power Meter Application Notes 6-4 Frequency Response. The carrier frequency response of a diode detector is determin

Página 150

Boonton 4540 Series RF Power Meter Application Notes 6-5 6.2 Pulse Definitions IEEE Std 194™-1977 Standard Pulse Terms and Definitions ―prov

Página 151

Boonton 4540 Series RF Power Meter Application Notes 6-6 Table 6-1. Pulse Terms TERM DEFINITION Base Line The two portions of a pulse w

Página 152

Boonton 4540 Series RF Power Meter Application Notes 6-7 6.3.2 Automatic Measurement Terms The following terms appear in the 4540 Series Tex

Página 153

Boonton 4540 Series RF Power Meter Application Notes 6-8 6.3.3 Automatic Measurement Sequence The automatic measurement process analyzes the captur

Página 154

Boonton 4540 Series RF Power Meter Application Notes 6-9 7. The process locates the bottom amplitude (baseline) using the IEEE histogram method. A h

Página 155

Boonton 4540 Series RF Power Meter Application Notes 6-10 Item dBm mW Mesial value 10.0 10.0 Sample n 8.0 6

Página 156

Boonton 4540 Series RF Power Meter Application Notes 6-11 12. The processor calculates the output values according to the following definitions: a

Página 157

Boonton 4540 Series RF Power Meter General Information 1-10 Table 1-2 4540 Series Performance Specifications (continued) (Specifications are subj

Página 158

Boonton 4540 Series RF Power Meter Application Notes 6-12 Figure 6-7 Sampling Intervals 14. The processor calculates the delay between

Página 159

Boonton 4540 Series RF Power Meter Application Notes 6-13 Fig. 6-9 Statistical Mode Text Display Page 1

Página 160

Boonton 4540 Series RF Power Meter Application Notes 6-14 Table 6-3. Statistical Automatic Measurements In the Statistical mode the following five a

Página 161

Boonton 4540 Series RF Power Meter Application Notes 6-15 The following two global status values are displayed: TEXT

Página 162

Boonton 4540 Series RF Power Meter Application Notes 6-16 6.5 Measurement Accuracy The 4540 Series includes a precision, internal, 50 MHz RF referen

Página 163

Boonton 4540 Series RF Power Meter Application Notes 6-17 The worst case approach is a very conservative method in which the extreme conditions of eac

Página 164

Boonton 4540 Series RF Power Meter Application Notes 6-18

Página 165

Boonton 4540 Series RF Power Meter Application Notes 6-19 The value to use for calibration level uncertainty depends upon the sensor calibration techn

Página 166

Boonton 4540 Series RF Power Meter Application Notes 6-20 Sensor Temperature Coefficient. This term is the error which occurs when the sensor’s tempe

Página 167

Boonton 4540 Series RF Power Meter Application Notes 6-21 The sensor calfactor uncertainty is due to uncertainties encountered while performing this f

Página 168

Boonton 4540 Series RF Power Meter Installation 2-1 2. Installation This section contains unpacking and repacking instructions, power requirements,

Página 169

Boonton 4540 Series RF Power Meter Application Notes 6-22 Step 4: The Source Mismatch Uncertainty is calculated using the formula in the

Página 170

Boonton 4540 Series RF Power Meter Application Notes 6-23 Step 10: Now that each of the individual uncertainty terms has been determined, we can combi

Página 171

Boonton 4540 Series RF Power Meter Application Notes 6-24 Step 2: The Calibrator Level Uncertainty for the Model 2530 External 1GHz calibr

Página 172

Boonton 4540 Series RF Power Meter Application Notes 6-25 Step 9: The Sensor Calfactor Uncertainty needs to be interpolated from the uncertainty valu

Página 174

Boonton 4540 Series RF Power Meter Appendix A 7-1 7. Maintenance This section presents procedures for maintaining the 4540 Series. 7.1 Safety

Página 175

Boonton 4540 Series RF Power Meter Appendix A 7-2 7.5 Firmware Upgrade Operating firmware has been loaded into the 4540 Series instrument at the fac

Página 176

Boonton 4540 Series RF Power Meter Appendix A 7-3 How to Start Up in Boot Loader Mode 1. From a powered-down state, press and hold the to

Página 177

Boonton 4540 Series RF Power Meter Appendix A 7-4 This page intentionally left blank.

Página 178

Boonton 4540 Series RF Power Meter Appendix A 8-5 8. Appendix A SCPI Error Messages 8.1 SCPI Error Messages NO. MESSAGE DESCRIPTIO

Página 179

Boonton 4540 Series RF Power Meter Installation 2-2 Table 2-1 4540 Series Packing List INSTRUMENT (See also Table 1-1) 4540 Series RF Power Meter

Página 180

Boonton 4540 Series RF Power Meter Appendix A 8-6 NO. MESSAGE DESCRIPTION -245 "Sensor Disconnected." -246 ―Sensor voltage e

Página 181

Boonton 4540 Series RF Power Meter Appendix A 8-7 NO. MESSAGE DESCRIPTION -342 "Temp comp pwr non-mono" -343 "Temp comp

Página 182

Boonton 4540 Series RF Power Meter Appendix A 8-8 This page intentionally left blank.

Página 183

Boonton 4540 Series RF Power Meter Appendix B 9-1 9. Appendix B Model 2530 1GHz Calibrator 9.1 Description The Model 2530 operates as an accesso

Página 184

Boonton 4540 Series RF Power Meter Appendix B 9-2 9.4 Specifications RF Characteristics Operating Modes: CW, Internal or external pulse Output Fre

Página 185

Boonton 4540 Series RF Power Meter Appendix B 9-3 9.5 Connections 1. Plug the supplied RJ-11 cable into the rear panel jack of the 2530 and into th

Página 186

Boonton 4540 Series RF Power Meter Appendix B 9-4 This page intentionally left blank.

Página 187

Boonton 4540 Series RF Power Meter Appendix C 10-1 10. Appendix C Boonton EULA "Boonton Electronics 4540 Series Operating Software" END

Página 188

Boonton 4540 Series RF Power Meter Appendix C 10-2 8. U.S. GOVERNMENT LICENSE RIGHTS. All Product provided to the U.S. Government pursuant

Página 189

Boonton 4540 Series RF Power Meter Appendix D 11-1 11. Appendix D Warranty & Repair Repair Policy 4540 Series Instrument. If the Boonton 4540

Página 190

Boonton 4540 Series RF Power Meter Installation 2-3 Trigger Most triggered applications can use the RF signal applied to the sensors for triggering.

Página 191

Boonton 4540 Series RF Power Meter Appendix D 11-2 THE FOREGOING WARRANTIES ARE IN LIEU OF ALL OTHER WARRANTIES, EXPRESS OR IMPLIED, INCLUD-ING, B

Página 192

Boonton 4540 Series RF Power Meter Installation 2-4 6. On the front panel, press the Menu key followed by the More softkey, then on the softkeys, pre

Página 193

Boonton 4540 Series RF Power Meter Getting Started 3-1 3. Getting Started This chapter will introduce the user to the 4540 Series. The chapter wil

Página 194

Boonton 4540 Series RF Power Meter Contents iii SAFETY SUMMARY The following general safety precautions must be observed during all phases of operat

Página 195

Boonton 4540 Series RF Power Meter Getting Started 3-2 Figure 3-1. Standard 4540 Series RF Power Meter - Front Panel Table 3-1 Operating Control

Página 196

Boonton 4540 Series RF Power Meter Getting Started 3-3 Table 3-1 Operating Controls, Indicators and Connections (continued) Reference # Front Rear

Página 197

Boonton 4540 Series RF Power Meter Getting Started 3-4 Table 3-1 Operating Controls, Indicators and Connections (continued) Reference # Front Rear

Página 198

Boonton 4540 Series RF Power Meter Getting Started 3-5 Table 3-1 Operating Controls, Indicators and Connections (continued) Reference # Front Rear

Página 199

Boonton 4540 Series RF Power Meter Getting Started 3-6 3.3 LCD Monitor Display This subsection includes a picture (Figure 3-3) and a table (Table 3-

Página 200

Boonton 4540 Series RF Power Meter Getting Started 3-7 Table 3-2 LCD Monitor Display Fields (continued) Ref. No. Field Name Mode Descriptio

Página 201

Boonton 4540 Series RF Power Meter Getting Started 3-8 Table 3-2 LCD Monitor Display Fields (continued) 10 GPIB/Remote Status All The GPIB/Re

Página 202

Boonton 4540 Series RF Power Meter Getting Started 3-9 Table 3-2 LCD Monitor Display Fields (continued) Ref. No. Field Name Mode Descriptio

Página 203

Boonton 4540 Series RF Power Meter Getting Started 3-10 3.4 Initialize The procedures presented in this section will initialize the 4540 Series and

Página 204

Boonton 4540 Series RF Power Meter Getting Started 3-11 Table 3-3. Initialized Parameters Parameter Applies to: Default Graph/Text Mod

Página 205

Boonton 4540 Series RF Power Meter Contents iv SAFETY SYMBOLS This safety requirement symbol (located on the rear panel) has been adopted by the Inte

Página 206

Boonton 4540 Series RF Power Meter Getting Started 3-12 Table 3-3. Initialized Parameters (continued) Parameter Applies to: Default Pa

Página 207

Boonton 4540 Series RF Power Meter Getting Started 3-13 Table 3-3. Initialized Parameters (continued) Parameter Applies to: Default Pa

Página 208

Boonton 4540 Series RF Power Meter Getting Started 3-14 3.5 Calibration Before any pulse measurements can be acquired with the 4540 Series, a peak p

Página 209

Boonton 4540 Series RF Power Meter Getting Started 3-15 STEP PROCEDURE 1. Select the Calibration menu soft key. Initiate autocalibration

Página 210

Boonton 4540 Series RF Power Meter Getting Started 3-16 Figure 3-5. Menus for Sensor Calibration Procedure

Página 211

Boonton 4540 Series RF Power Meter Getting Started 3-17 3.6 Making Measurements What you need to know To perform accurate measurements, the followin

Página 212

Boonton 4540 Series RF Power Meter Getting Started 3-18 3.6.2 Pulse Mode For periodic or pulsed signals, it is often necessary to analyze the power f

Página 213

Boonton 4540 Series RF Power Meter Getting Started 3-19 3.6.3 Statistical Mode Certain signals are completely random and provide no event that can se

Página 214

Boonton 4540 Series RF Power Meter Getting Started 3-20 Note that this analysis does not depend upon any particular test signal, or upon synchronizati

Página 215

Boonton 4540 Series RF Power Meter Operation 4-1 4. Operation This section presents the control menus and procedures for operating the 4540 Series i

Página 216

Boonton 4540 Series RF Power Meter Contents v 1. General Information ...

Página 217

Boonton 4540 Series RF Power Meter Operation 4-2 4.3 Numerical Data Entry Many 4540 Series parameters can be entered directly from the numeric keypa

Página 218

Boonton 4540 Series RF Power Meter Operation 4-3 4.4 Sensor Calibration The Model 4540 Series features a built-in AutoCal process that automa

Página 219

Boonton 4540 Series RF Power Meter Operation 4-4 4.5 Menu Reference 4.5.1 Main Menu - Pulse and Modulated Modes - Pg 1 (Top Level) The Main Menu i

Página 220

Boonton 4540 Series RF Power Meter Operation 4-5 4.5.2 Main Menu - Pulse and Modulated Modes - Pg 2 (Top Level) The Main Menu is the topmost menu lev

Página 221

Boonton 4540 Series RF Power Meter Operation 4-6 4.5.3 Main Menu - Statistical Mode - Pg 1 (Top Level) The Main Menu is the topmost menu level from

Página 222

Boonton 4540 Series RF Power Meter Operation 4-7 4.5.4 Main Menu - Statistical Mode - Pg 2 (Top Level) The Main Menu is the topmost menu level from w

Página 223

Boonton 4540 Series RF Power Meter Operation 4-8 4.5.5 Measure Menu The Measure menu contains items for controlling measurement acquisition and mode.

Página 224

Boonton 4540 Series RF Power Meter Operation 4-9 Main > Measure > Clear Description: Clear display traces and all data buffers for CH1 and CH2.

Página 225

Boonton 4540 Series RF Power Meter Operation 4-10 4.5.6 Channel Menu Select the specific menu for the desired measurement or reference channel if ava

Página 226

Boonton 4540 Series RF Power Meter Operation 4-11 4.5.7 Channel 1 | 2 Menu The Channel menu contains items that affect a single measurement

Página 227

Boonton 4540 Series RF Power Meter Contents vi 4. Operation ...

Página 228

Boonton 4540 Series RF Power Meter Operation 4-12 Main > Channel > Channel 1 | 2 > Calibration Description: Select the calibration menu for

Página 229 - 6. Application Notes

Boonton 4540 Series RF Power Meter Operation 4-13 4.5.8 Channel > Calibration Menu The calibration menu contains menu items to initiate multi-step

Página 230 - Application Notes

Boonton 4540 Series RF Power Meter Operation 4-14 Main > Channel > Channel 1 | 2 > Calibration > Sel Calibrator Description: Selects the

Página 231

Boonton 4540 Series RF Power Meter Operation 4-15 4.5.9 Channel > Extensions Pg 1 Menu (Modulated Mode) The extensions menu provides access

Página 232

Boonton 4540 Series RF Power Meter Operation 4-16 Main > Channel > Channel 1 | 2 > Extensions Pg 1 > Filter Time Description: Set the c

Página 233 - 6.2 Pulse Definitions

Boonton 4540 Series RF Power Meter Operation 4-17 4.5.10 Channel > Extensions Pg 1 > Corrections Menu (Modulated Mode) The corrections menu co

Página 234 - 6.3 Automatic Measurements

Boonton 4540 Series RF Power Meter Operation 4-18 F

Página 235

Boonton 4540 Series RF Power Meter Operation 4-19 4.5.11 Channel > Extensions Pg 2 Menu (Modulated Mode) This menu is page two of the channel exte

Página 236

Boonton 4540 Series RF Power Meter Operation 4-20 Main > Channel > Channel 1 | 2 > Extensions Pg 2 > Frequency Description: Set the RF f

Página 237

Boonton 4540 Series RF Power Meter Operation 4-21 4.5.12 Channel > Extensions Pg 2 > Alarms Menu (Modulated Mode) The Alarms menu provides co

Página 238

Boonton 4540 Series RF Power Meter Contents vii 4.5.39 System > I/O Config > Multi IO Menu ...

Página 239

Boonton 4540 Series RF Power Meter Operation 4-22 Main > Channel > Channel 1 | 2 > Extensions Pg 2 > Alarms > Low Limit Description:

Página 240

Boonton 4540 Series RF Power Meter Operation 4-23 4.5.13 Channel > Extensions Pg 1 Menu (Pulse Mode) The extensions menu provides access to

Página 241

Boonton 4540 Series RF Power Meter Operation 4-24 Main > Channel > Channel 1 | 2 > Extensions Pg 1 > Define Pulse Description: Selects t

Página 242

Boonton 4540 Series RF Power Meter Operation 4-25 4.5.14 Channel > Extensions Pg 1 > Corrections Menu (Pulse Mode) The corrections menu contai

Página 243

Boonton 4540 Series RF Power Meter Operation 4-26 Main > Channel > Channel 1 | 2 > Extensions Pg 1 > Corrections > Duty Cycle Descript

Página 244 - 6.5 Measurement Accuracy

Boonton 4540 Series RF Power Meter Operation 4-27 4.5.15 Channel > Extensions Pg 1 > Define Pulse Menu (Pulse Mode) The define pulse menu conta

Página 245

Boonton 4540 Series RF Power Meter Operation 4-28 Main > Channel > Channel 1 | 2 > Extensions Pg 1 > Define Pulse > Pulse Units Descrip

Página 246

Boonton 4540 Series RF Power Meter Operation 4-29 4.5.16 Channel > Extensions Pg 1 > Define Pulse > Pulse Gate Menu Main Channel > Chan

Página 247

Boonton 4540 Series RF Power Meter Operation 4-30 4.5.17 Channel > Extensions Pg 2 Menu (Pulse Mode) This menu is page two of the channel extensi

Página 248

Boonton 4540 Series RF Power Meter Operation 4-31 Main > Channel > Channel 1 | 2 > Extensions Pg 2 > Frequency Description: Set the RF fr

Página 249

Boonton 4540 Series RF Power Meter Contents viii *SRE ...

Página 250

Boonton 4540 Series RF Power Meter Operation 4-32 4.5.18 Channel > Extensions Pg 2 > Alarms Menu (Pulse Mode) The Alarms menu provides contro

Página 251

Boonton 4540 Series RF Power Meter Operation 4-33 Main > Channel > Channel 1 | 2 > Extensions Pg 2 > Alarms > Low Limit Description:

Página 252

Boonton 4540 Series RF Power Meter Operation 4-34 4.5.19 Channel > Extensions Pg 1 Menu (Statistical Mode) The extensions menu provides access t

Página 253

Boonton 4540 Series RF Power Meter Operation 4-35 4.5.20 Channel > Extensions Pg 1 > Corrections Menu (Statistical Mode) The corrections menu

Página 254

Boonton 4540 Series RF Power Meter Operation 4-36 Main > Channel > Channel 1 | 2 > Extensions Pg 1 > Corrections > Duty Cycle Descript

Página 255 - 7. Maintenance

Boonton 4540 Series RF Power Meter Operation 4-37 4.5.21 Channel > Extensions Pg 2 Menu (Statistical Mode) This menu is page two of the channel ex

Página 256 - 7.5 Firmware Upgrade

Boonton 4540 Series RF Power Meter Operation 4-38 Main > Channel > Channel 1 | 2 > Extensions Pg 2 > Frequency Description: Set the RF f

Página 257

Boonton 4540 Series RF Power Meter Operation 4-39 4.5.22 Channel > Extensions Pg 2 > Alarms Menu (Statistical Mode) The Alarms menu provides

Página 258

Boonton 4540 Series RF Power Meter Operation 4-40 Main > Channel > Channel 1 | 2 > Extensions Pg2 > Alarms > Low Limit Description:

Página 259 - 8.1 SCPI Error Messages

Boonton 4540 Series RF Power Meter Operation 4-41 4.5.23 Channel > Ref 1 | 2 Menu The Reference Menu is used to manually or automatically set a p

Página 260

Boonton 4540 Series RF Power Meter Contents ix 5.6.5 FETCh Queries ...

Página 261

Boonton 4540 Series RF Power Meter Operation 4-42 Main > Channel > Ref 1 | 2 > Set Ref Description: Manually set the ratiometric mode ref

Página 262

Boonton 4540 Series RF Power Meter Operation 4-43 4.5.24 Trigger Menu (Pulse Mode) The Trigger menu is used to configure trigger settings for pulse m

Página 263 - 9.3 Optional Configurations

Boonton 4540 Series RF Power Meter Operation 4-44 Main > Trigger > Trig Slope Description: Set the trigger slope or polarity. When set to +, t

Página 264 - 9.4 Specifications

Boonton 4540 Series RF Power Meter Operation 4-45 4.5.25 Time Menu (Modulated and Pulse Mode) The time menu provides for controlling the timebase set

Página 265 - 9.5 Connections

Boonton 4540 Series RF Power Meter Operation 4-46 Main > Time > Trig Delay Description: Set the trigger delay time in seconds with respect to

Página 266

Boonton 4540 Series RF Power Meter Operation 4-47 4.5.26 Stat Mode Menu (Statistical Mode) The statistical mode menu provides for controlling the

Página 267 - 10. Appendix C Boonton EULA

Boonton 4540 Series RF Power Meter Operation 4-48 4.5.27 Term Options Menu (Statistical Mode) The statistical mode menu provides for controlling the

Página 268 - Appendix C

Boonton 4540 Series RF Power Meter Operation 4-49 4.5.28 Cursors Menu (Statistical Mode) The statistical cursors menu controls the mode and the verti

Página 269 - Limited Warranty

Boonton 4540 Series RF Power Meter Operation 4-50 4.5.29 Markers Menu (Modulated and Pulse Modes) The Markers menu is used to configure measurement m

Página 270

Boonton 4540 Series RF Power Meter Operation 4-51 4.5.30 Display Menu The Display menu is the entry point for accessing and editing display configura

Comentários a estes Manuais