Remote Operation
5-42
OUTPut:[MIO:]RECOroderMIN
Description: Set or return the Multi IO recorder output minimum, or downscale (-10.0V or 0.0V) power
reference level. For voltage probes, this is the equivalent power at the current impedance setting.
Syntax: OUTPut:RECOrder:MIN < numeric_value >
Argument: < numeric_value > = -200.00 to +200.00 dBm
Valid Modes: Any
OUTPut:[MIO:]RECOroder:POLarity
Description: Set or return the recorder output polarity. UNIPOLAR selects 0.0 volts and 10.0 volts to as
minimum and maximum output levels, and BIPOLAR selects -10.0 to +10.0 volts.
Syntax: OUTPut:RECOrder:POLarity <character data>
Argument: <character data> = UNIPOLAR, BIPOLAR
Valid Modes: Any
OUTPut:[MIO:]RECOroder:SCALing
Description: Set or return the Multi IO measurement mode when in recorder mode. If set to AUTO, the output
level is automatically scaled to match the display, and generally will ―downscale‖ every decade as
power increases. In MANUAL mode, the output level is scaled using the minimum and maximum
powers set with OUTPut:[MIO:]RECOrder:MIN and :MAX as the downscale and full scale
values. This allows very wide range or very high resolution.
Syntax: OUTPut:[MIO:]RECOrder:SCAling <character data>
Argument: <character data> = AUTO, MANUAL
Valid Modes: All
OUTPut:[MIO:]RECOroder:SOURce
Description: Set or return the source channel for the Multi IO recorder output.
Syntax: OUTPut:[MIO:]RECOrder:SOURce <character data>
Argument: <character data> = CH1, CH2
Valid Modes: All