Boonton 4540 Peak Power Meter User Manual Manual do Utilizador Página 173

  • Descarregar
  • Adicionar aos meus manuais
  • Imprimir
  • Página
    / 270
  • Índice
  • MARCADORES
  • Avaliado. / 5. Com base em avaliações de clientes
Vista de página 172
Boonton 4540 Series RF Power Meter
Remote Operation
5-39
OUTPut:EXTernal:PRESent?
Description: Return true if the Model 2530 external 1 GHz calibrator is connected and powered up. Otherwise,
return false.
Syntax: OUTPut:EXTernal:PRESent? <Boolean>
Returns: <Boolean> = { 0, 1 }, Accepts { 0, 1, ON, OFF }
Valid Modes: Any
OUTPut:EXTernal:PULSe:CTRL
Description: Set or return the pulse control mode of the external 1 GHz calibrator’s pulse modulator. This
parameter has no effect in the External source pulse control mode.
Syntax: OUTPut:EXTernal:PULSe:CTRL <character data>
Argument: <character data> = {PRESET, VAR}
Valid Modes: Any
OUTPut:EXTernal:PULSe:DCYCle
Description: Set or return the duty cycle in percent in the Preset pulse control mode of the external 1 GHz
calibrator. This parameter has no effect in the variable pulse control mode or the external pulse
source mode. Note: If the argument value is greater than 50% OUTPut:EXTernal:POLArity is
forced to NEG, otherwise POS.
Syntax: OUTPut:EXTernal:PULSe:DCYCle <numeric_value>
Argument: <numeric_value> = 10% to 90% in 10% increments
Valid Modes: Any
OUTPut:EXTernal:PULse:PERiod
Description: Set or return the pulse period in the preset pulse control mode of the external 1 GHz calibrator.
This parameter has no effect in the variable pulse control mode external pulse source mode.
Syntax: OUTPut:EXTernal:PULse:PERiod <numeric_value>
Argument: <numeric_value> = {10e-3, 1e-3, 100e-6} seconds, or
{10ms, 1ms, 0.1ms}, or
{10000us, 1000us, 100us}
Valid Modes: 1 GHz Calibrator Preset Pulse control Mode
Vista de página 172
1 2 ... 168 169 170 171 172 173 174 175 176 177 178 ... 269 270

Comentários a estes Manuais

Sem comentários