Boonton 4240 Series RF Power Meter
STATus:QUEStionable:CONDition?
f the Questionable Condition register. The following table shows the
Bit Va
Description: Return the current value o
bit assignments in the register. These bits are updated by the instrument in real time, and can
change in response to changes in the instrument’s operating condition.
lue Definition
1 Not used always 0.
2 Not used always 0.
0
1
invalid.
1 always 0.
3 always 0.
6 always 0.
7 128 Not used always 0.
8 256 Calibration 1 = sensor requires calibration and/or zeroing
9 512 Not used always 0.
always 0.
11 2048 Not used always 0.
40
13 81
14 16
15 32
Status summary bit will be set.
yntax: STATus:QUEStionable:EVENt?
2 4 Not used always 0.
3 8 Power 1 = a power measurement may be
4 6 Not used
5 2 Not used
6 4 Not used
10 1024 Not used
12 96 Not used always 0.
92 Not used always 0.
384 Not used always 0.
768 Not used always 0.
Syntax: STATus:QUEStionable:CONDition?
Returns: 16-bit register value (0 to 65535)
S
TATus:QUEStionable:ENABle
Description: Sets or returns the Questionable Enable register, which contains the bit mask that defines which
true conditions in the Questionable Status Event register will be reported in the Questionable
Summary bit of the instrument Status Byte. If any bit is 1 in the Questionable Enable register and
its corresponding Questionable Event bit is true, the Questionable
Syntax: STATus:QUEStionable:ENABle <numeric value>
Argument: <numeric value> = 0 to 65535
Returns: 16-bit register value (0 to 65535)
TATus:QUEStionable:EVENt? S
Description: Returns the current contents of the Questionable Event register then resets the register value to 0.
The Questionable Event register contains the latched events from the Questionable Condition
register as specified by the Questionable status group’s positive and negative transition filters.
S
Returns: 16-bit register value (0 to 65535)
Remote Operation