Boonton 4240 Series RF Power Meter
Table 1-2 4240 Series Performance Specifications (continued)
pecifications a
NCERTAINTIES
easurement Accuracy Sum of following uncertainties (errors are +
(S re subject to change without notice)
U
M worst case): instrument
uncertainty, noise/signal percentage, power reference uncertainty,
s , temperature drift, mismatch, and frequency
c
Instrume
+
ensor shaping
alibration factors
nt Uncertainty
0.23% (+0.01 dB) at full scale; +0.46% (+0.02 dB) at 1/10 full scale
Noise/si C wer Sensor Manual, to percent
power level.
Power Reference Uncertainty R nce: Level Accuracy
Sensor Shaping
+
gnal Percentage onvert 2 sigma noise listed in the Po
of the applied
efer to Table 1-2 Power Refere
1.0% (+0.04 dB) typical, Power Sensor Manual
emperature Drift R
MEASUREMENT SYSTEM
Sensor inputs: One or two sensor measurement channels.
Measurement Technique: 24 bit Sigma-delta A/D converter per channel.
CALIBRATION SOURCE
Internal Calibrator
Operating Modes: Off, On CW
Frequency: 50.025 MHz ± 0.1%
Level Range: -60 to +20 dBm
Resolution: 0.1 dB
RF Connector: Type N
Source VSWR: 1.05 (reflection coefficient = 0.024)
Accuracy, 0C to 20C, NIST traceable:
At 0 dBm: ±0.055 dB (1.27%)
+20 to -39 dBm: ±0.075 dB (1.74%)
-40 to -60 dBm: ±0.105 dB (2.45%)
Auto-calibration: The Calibrator is used to automatically generate linearity calibration
data for CW power sensors. Can be used to provide test signals.
efer to Power Sensor Manual T