PIM31-F07-UL LTE Analyzer Mode:
The PIM31 LTE Analyzer software enables the user to analyze PIM
measurements in greater depth due to its three modes of operation:
Frequency
Time
Sweep
Element
Name
Description / Display
A
Pull-Down menu
Allows screenshots, data logging and self test. For
more information refer to the Pull-Down Menus
chapter.
B
Parameter Setting
for Carrier 1
Allows modification of frequencies and power levels
used in Analyzer Time and Frequency modes.
Works only with RF Power OFF.
C
Parameter Setting
for Carrier 2
Allows modification of frequencies and power levels
used in Analyzer Time and Frequency modes.
Works only with RF Power OFF.
D
Graphical display
for Upper Band
Shows up to 2 highest IM signals for upper band as
a vertical bar. Position on the x-axis indicates
frequency, and length indicates power level. Color
of the bars refers to the colors of the numeric
displays (O-P).
Vista de página 104